ASAP 2020

logo_without_tag

 
ASAP 2020

 


Description

Micromeritics’ ASAP 2020 Accelerated Surface Area and Porosimetry analyzer uses the gas sorption technique to generate high-quality data for research and quality control applications. Available options include the micropore option, the high-vac option, and the chemisorption option, which uses the static volumetric technique to determine the percent metal dispersion, active metal surface area, size of active particles, and surface acidity of catalyst materials. Data obtained by an ASAP 2020 analyzer are used to achieve efficiency, safety, quality, and profitability in any modern materials-oriented work, and can be used by scientists with confidence and assurance.

Uses of the ASAP 2020
The ASAP 2020 is used to monitor catalysts converting a stream of feedstock chemicals into a product, activated carbon used for the recovery of precious metals from a mining process solution, a pharmaceutical tablet as it undergoes dissolution and absorption at just the right dosing rate, and production of ceramics made from an exacting combination of finely powdered raw materials. Micromeritcs’ staff of scientists and engineers can help you determine how the ASAP 2020 can be used for your application.

 


Features

  • Two Independent vacuum systems that allow preparation and analysis to proceed concurrently
  • Oil-free “dry” vacuum option to prevent oil contamination
  • Two-station intelligent degas system for fully automated degassing with controlled heating time profiles
  • long-duration cryogen system for unattended analyses
  • Twelve gas inlets that are automatically selectable to allow for automated selection of pretreatment, backfill, and analysis gases, and the ability to connect to a mass spectrometer

Analysis Technique

The basics of the analytical technique are simple; a sample contained in an evacuated sample tube is cooled (typically) to cryogenic temperature, then is exposed to analysis gas at a series of precisely controlled pressures. With each incremental pressure increase, the number of gas molecules adsorbed on the surface increases. The pressure at which adsorption equilibrium occurs is measured and the universal gas law is applied to determine the quantity of gas adsorbed.

As adsorption proceeds, the thickness of the adsorbed film increases. Any micropores in the surface are quickly filled, then the free surface becomes completely covered, and finally larger pores are filled. The process may continue to the point of bulk condensation of the analysis gas. Then, the desorption process may begin in which pressure systematically is reduced resulting in liberation of the adsorbed molecules. As with the adsorption process, the changing quantity of gas on the solid surface is quantified. These two sets of data describe the adsorption and desorption isotherms. Analysis of the isotherms yields information about the surface characteristics of the material.

 


Specification

Pressure Measurement: Range: 0 to 950 mmHg
Resolution:
  • 1000-mmHg Transducer: 0.001 mmHg
  • 10-mmHg Transducer*: 0.00001 mmHg
  • 1-mmHg Transducer**: 0.000001 mmHg (optional for expanded resolution)
Accuracy (Analysis System Only):
Includes nonlinearity, hysteresis, and nonrepeatability. Transducer manufacturer’s specifications.
  • 1000-mmHg Range: within 0.15% of reading (analysis system only)
  • 10-mmHg Range*: within 0.15% of reading
  • 1-mmHg Range**: within 0.12% of reading
System Capacity: Sample Preparation: 2 integrated degas ports
Analysis: 1 sample port and 1 saturation pressure tube
Up to 2 complete analysis units can be independently controlled by 1 computer
Vacuum System: Degas: Either two-stage, oil sealed, mechanical vacuum pump, 5 x 10-3 mmHg ultimate vacuum or high vacuum system ultimate vacuum <10-8 mbar
Analysis – Nitrogen system: two stage, oil sealed mechanical vacuum pump, 5 x 10-3 mmHg ultimate vacuum.
Multigas and Micropore: High vac system, ultimate <10-8 mbar.
Degas System: Temperature Range: Ambient to 450 ºC(Programmable)
Temperature Selection: Digitally set, 1 ºC increments
Accuracy: Deviation less than +/- 10 ºC of set point at thermocouple
Backfill Gas: User-selectable, typically nitrogen or helium
Cryogen System: Patented Isothermal Jackets effectively maintain cryogen level constant on sample tube and P0 tube during analysis as evaporation of cryogen occurs.
Capacity: 3-Liter Dewar, which provides up to 72 hours of unattended analysis.
Analysis Time: Unlimited. Cryogen Dewars may be refilled without affecting the accuracy of results
Sample Size: Sample tubes are available for various size pellets, cores, and powders. Sample tube stems are 1/2-in.(1.27-cm) OD with 9-cm3 bulbs, 1/4-in. (0.635-cm) or 3/8-in. (0.953-cm) OD stems with 9-cm3 bulbs. Special tubes can be designed.
Electrical: Voltage: 100, 115, 230 VAC +/-10%
Frequency: 50/60 Hz
Power: 700 VA, operating (maximum)
Physical: Height: 99 cm (39 in.)
Width: 85 cm (33.5 in.)
Depth: 61 cm (24 in.)
Weight: 115 kg (250 lbs)
Miscellaneous: ISO 9001 manufacturer , CE certified

 

 

                                                                                                                                                                                                                                                                                                                                                                                                                           